.Design and experiment of a laser ionization source for ion mobility spectrometer[J].Optoelectronics Letters,2006,2(5):399-402
Design and experiment of a laser ionization source for ion mobility spectrometer
JI .Ren-dong KONG Xiang-he LIU Xian-yun ZHANG Shu-dong
College of Physics and Engineering, Qufu Normal University, Qufu 273165. China
Abstract:
      The fabrication and characterization of a laser ionization source for ion mobility spectrometer (IMS) is described with the calculated parameter of the drift tube. lons are created in alr at ambient pressure using laser ionization (λ=248 nm) with air as the drift gas. Mobility spectra for positive ions at different electric fields are obtained with the length of drift tube of 25 and 50 nm, respectively. The distribution of the electric field in the drift tube (50 nm) has also been investigated by SIMION simulation with the ion shutter closed and open respectively.
Hits: 3596
Download times: 182
View Full Text    Download reader