.Accurate measurement of magnesium content in alpha-olefins by laser induced breakdown spectroscopy (LIBS) technique[J].Optoelectronics Letters,2007,3(3):222-226 |
Accurate measurement of magnesium content in alpha-olefins by laser induced breakdown spectroscopy (LIBS) technique |
U.Visentini G. Cristoforetti S. Legnaioli A. Salvetti E. Tognoni V. Palleschi |
[1]Basell R &D, Polymer Physics and Characterization, Piazzale Donegani, 12 QCL 44100 Ferrara, Italy [2]Applied Laser Spectroscopy Laboratory, Istituto per i Processi Chimico-Fisici, Area di Ricerca del Consiglio Nazionale delle Ricerche, Via Moruzzi 1, 56124 Pisa, Italy |
Abstract: |
In this paper, results are presented on first application of laser-induced breakdown spectroscopy (LIBS) technique for quantitative determination of Mg concentration in alpha-olefins. It is demonstrated that this technique gives accuracies in Mg concentrations which are comparable with those of traditional methods, as ICP-OES, in short times and without need for sample preparation. Limits of detection for Mg of the order of 500 ppb are demonstrated. Moreover, a brief discussion about the realization of proper calibration curves and the definition of the limit of detection (LOD) is reported. |
Hits: 3557 |
Download times: 191 |
|
View Full Text Download reader |
|
|
|